Recommendations For Standards of Monitoring and Safety during


Download Recommendations For Standards of Monitoring and Safety during


Preview text

Recommendations
For Standards of
Monitoring and Safety during
Cardiopulmonary Bypass
Published by the Society of Clinical Perfusion Scientists of Great Britain
& Ireland Association for Cardiothoracic Anaesthesia and
Critical Care Society for Cardiothoracic Surgery in Great Britain
& Ireland August 2016

Recommendations for Standards of Monitoring and Safety during Cardiopulmonary Bypass (CPB)

Membership of the working party (2007)

Ian Curle

Society of Clinical Perfusion Scientists of Great Britain & Ireland

Fiona Gibson

Association of Cardiothoracic Anaesthetists

Jonathan Hyde

Society for Cardiothoracic Surgery in Great Britain and Ireland

Alex Shipolini

Society for Cardiothoracic Surgery in Great Britain & Ireland

David Smith

Association of Cardiothoracic Anaesthetists

J P van Besouw Association of Cardiothoracic Anaesthetists

Kate Wark

Association of Cardiothoracic Anaesthetists

Mike Weatherall

Society of Clinical Perfusion Scientists of Great Britain & Ireland

David Whitaker Association of Cardiothoracic Anaesthetists

Revision Personnel (2016)

Andrew Heggie

Society of Clinical Perfusion Scientists of Great Britain & Ireland

Simon Anderson

Society of Clinical Perfusion Scientists of Great Britain & Ireland

Thassee Pillay

Society for Cardiothoracic Surgery in Great Britain and Ireland

Heyman Luckraz

Society for Cardiothoracic Surgery in Great Britain and Ireland

Niall O’Keeffe

Association for Cardiothoracic Anaesthesia and Critical Care

Moyna Bill

Association for Cardiothoracic Anaesthesia and Critical Care

This document is available on the following websites: Society of Clinical Perfusion Scientists of Great Britain & Ireland www.scps.org.uk Association for Cardiothoracic Anaesthesia and Critical Care www.acta.org.uk Society for Cardiothoracic Surgery in Great Britain & Ireland www.scts.org

Introduction
The aim of this document is to determine standards of monitoring and safety during CPB for adult and paediatric surgery. These standards are considered by the Society of Clinical Perfusion Scientists of Great Britain and Ireland, the Association for Cardiothoracic Anaesthesia and Critical Care and the Society for Cardiothoracic Surgery in Great Britain and Ireland to be the minimal monitoring required during CPB. This includes monitoring for the onset of and weaning from CPB, and for confirmation of anticoagulation and ventilation of the lungs.
These standards are for use in conjunction with the Society of Clinical Perfusion Scientists of Great Britain and Ireland Standards of Practice document1 and local protocols. Sources of reference include publications from the Society of Clinical Perfusion Scientists of Great Britain and Ireland1, the Association of Anaesthetists of Great Britain and Ireland2, the American Society of Extra-Corporeal Technology3 and the DH document – Guide to Good Practice in Clinical Perfusion (GGPCP)4.
(Within this document “on site facility” is defined as on the hospital site, “near patient facility” is defined as within or in close proximity to the cardiac theatre).
All centres undertaking cardiac surgery involving CPB should plan to institute these standards by 6 months from the date of publication.
The safe conduct of CPB is the joint responsibility of surgeons, anaesthetists and clinical perfusionists and requires a high level of communication between the team members. Whilst it is considered best practice during the conduct of CPB for a Consultant Cardiac Surgeon and Anaesthetist to be present in the operating room at all times during cardiopulmonary bypass, it is recognised that there are circumstances where this may not be essential. However, the safety

of CPB remains the primary responsibility of the perfusionist who must be present at all times. These situations should be managed using locally agreed clinical governance guidelines and protocols should be in place to ensure patient safety.
(Examples of such locally agreed guidelines, in relation to anaesthetic practice, can be found on the ACTA/ACACC website.)
Only an accredited clinical perfusionist registered with the College of Clinical Perfusion Scientists of Great Britain and Ireland must undertake or supervise the conduct of CPB1, 4, 5&6. A named and accredited clinical perfusionist without the distraction of other clinical commitments, in close proximity and freely available must supervise a trainee undertaking a CPB1.
This document should be used in conjunction with the World Health Organisation (WHO) surgical safety checklist or similar (e.g. Scottish Patient Safety Programme) and the Five Steps to Safer Surgery7 (or equivalent), and as guidance with these standards according to local protocols and policy.
This document is a review of the original guidelines—it will be reviewed regularly and may be revised or updated before the formal publication of a new edition. For the latest version, please visit the previously referred to websites.

General Recommendations
Adequate records of monitoring should be kept at all times and where a variable is monitored it should be regularly recorded. All units should have electronic acquisition and storage of this data, including the ability to produce a printout2. Records should be stored electronically and retained in the patient’s notes with this record containing all the variables monitored regularly during CPB.
All monitors and alarms used should be calibrated and maintained regularly according to the manufacturer’s instruction and the recommended service schedule. All equipment must be checked before use.
During CPB, the electrocardiogram (ECG), intravascular pressures and core body temperature should be continuously displayed and clearly visible to the clinical perfusionist, surgeon and anaesthetist. (Ideally, three separate screens).
Patient Specific Directions (PSD) / Patient Group Directives (Wales) should be in place as agreed at local level and, where not forming a part of a CPB record, should be a document recording those directions for each individual patient and according to guidance in the GGPCP4.
All departments should be familiar with the GGPCP4 and this should be readily available for reference.
All departmental team members should receive adequate training on any device or equipment to be used with the Manufacturer’s Instructions for Use (I.F.U.s) forming an integral part of that training and a record of training kept accordingly.

Monitoring of clinical parameters acquired directly from the patient
The following should be monitored and recorded continuously, (with local protocols dictating the frequency of recording):-
• Electrocardiogram (ECG)
• Systemic arterial pressure
• Central venous pressure
• Core body temperature
• Urine output should be monitored using a freely draining urinary catheter.
• Pulse oximetry should be continuously displayed when there is a spontaneous pulsatile circulation.
• Expired carbon dioxide tension/concentration should be continuously displayed when the lungs are being ventilated.
• The use of Bispectral Index (BIS) monitoring should be considered particularly in the presence and use of volatile Anaesthetic agents during CPB.

• The use of Near InfraRed Spectroscopy (NIRS) for Cerebral Oximetry monitoring should also be considered for certain procedures requiring CPB, e.g. Paediatric procedures, Aortic Root / Arch surgery, in patients with Carotid artery disease, procedures involving the use of retrograde arterial flow etc. (This list is not exhaustive and intended only as a guide).
N.B. It is accepted that special clinical circumstances (for example emergency surgery or failure to insert a urinary catheter) may, on some occasions, preclude complete monitoring.

Monitoring associated with the CPB circuit
The following should be monitored continuously and recorded accordingly:-
• Oxygen saturation of the blood in the venous return line.
• Oxygen tension or saturation of the blood in the arterial line.
• Continuity of the fresh gas flow to the oxygenator using an in-line flow meter or rotameter2.
• Oxygen concentration in the gas circuit to the oxygenator using an oxygen analyser with alarms, sited after the oxygen blender and vaporiser if used 2.
• Blood flow rate generated by the arterial pump.
• Arterial line pressure with consideration given to monitoring and recording the pressure pre-membrane pressure in addition to that of post-membrane.
• Cardioplegia delivery line pressure & temperature when Cardioplegia is delivered using the heart lung machine.

• Temperature of the blood in the venous & arterial limbs of the CPB circuit.
• Temperature of water in the heater/cooler system.
• Fluid record; all fluids, drugs and blood products added to the extra corporeal circuit pre & peri-CPB should be checked as required and, in the absence of a separate PSD, recorded on the perfusion chart and signed for accordingly7. Filtrate volume should be measured and recorded when a Haemofilter/ concentrator are being used.
• Continuous “in line” monitoring screens (e.g. for Haemoglobin / Hct, arterial & venous O2 saturation and / or tension, CO2 tension etc.) should be available and used when considered necessary.
The following measurements should be available at a near patient facility, (with local protocols dictating the frequency of recording):-
• Anticoagulation should be confirmed by an acceptable method e.g. Activated Clotting Time (ACT), which should be measured after heparinisation and before CPB and should be measured at regular intervals during CPB to ensure adequate anticoagulation.
Other parameters that should be measured are:-
• Blood gases

Preparing to load PDF file. please wait...